OPERATION
227
O
Interface
LOAD
GPIB
GPIB
05/04/22
16 : 50
Address
01
Configure
Other
Interface
System
Info
Load
Time Set
7.
Use the selector knob to edit the
GPIB address.
ENTER
or
8.
Edit the GPIB address.
Range
1 ~ 30
9.
Connect the GPIB cable to the
rear panel port: 24-pin female
connector.
GPIB constraints
Maximum 15 devices altogether, 20m cable
length, 2m between each device
Unique address assigned to each device
At least 2/3 of the devices turned On
No loop or parallel connection
Note
The GPIB Address must match that of the host
machine.
For GPIB function check, please refer to the section
“GPIB function check” on page 231.