![GW Instek PEL-2000B Series User Manual Download Page 146](http://html1.mh-extra.com/html/gw-instek/pel-2000b-series/pel-2000b-series_user-manual_2280092146.webp)
PEL-2000B Series User Manual
146
Creating a Program Sequence
Background
The PEL-2000B series has a total of 12 different
programs and 10 sequences to each program. That
totals to 120 different configurations.
Each Sequence in each program uses the settings
saved from Memory Data (Memory MXXX).
Memory Data contains settings such as the mode
and range for each channel. Different Sequences
can use the same Memory Data repeatedly. Each
Sequence loads all channels at the same time,
unless programmed otherwise.
CH2 M001
CH3 M001
CH1 M001
Run
On-Time
Off-Time
Short-Time
Short CH1
Short CH8
~
P/F-Time
CH5 M001
CH6 M001
CH4 M001
CH8 M001
CH7 M001
Sequence1
Sequences are run sequentially to create a
Program. There are 10 Sequences in each Program.
1
2
3
4
5
Sequence
Program 1
(P01)
6
7
8
9
10
If less than 10 Sequences are desired for a
Program, any additional Sequences can be
skipped (not run).
Sequence 2 & 3 are skipped.
1
2
3
4
5
Sequence
Program 1
(P01)
6
7
8
9
10
Summary of Contents for PEL-2000B Series
Page 1: ...Programmable DC Electronic Load PEL 2000B Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Page 5: ...Table of Contents 5 INDEX 312...
Page 113: ...TUTORIALS 113 14 Sound Turn the sound on off for the mainframe IU Page 202...
Page 309: ...APPENDIX 309 PEL 2004B 435 4 177 532 3 195 3 481 8 554 1 533 3 583 3...
Page 310: ...PEL 2000B Series User Manual 310 PEL 2020B PEL 2030B PEL 2040B PEL 2041B 81 6 171 2 475 446 8...
Page 317: ...INDEX 317 Wiring procedure 46...