OPERATION
191
3.
Press the Selector knob to edit the
selected setting, turn to change
the setting.
OR
Use the number pad to enter a
number.
9
P9
6
P6
P3
CLEAR
LOCK
3
8
P8
5
P5
P2
CAL.
2
7
P7
4
P4
P1
0
P0
1
CCP Step
0.002
A
4.
Press the Selector knob or Enter
to confirm selection.
ENTER
or
5.
Repeat steps 2-4 to edit the step resolution of:
CCL Step
CRH Step
CRL Step
CVH Step
CVL Step
CPH Step
CPL Step
Note
The Step resolution settings only apply to the active
channel, other channels will not be affected.
Configuring Response Time
Background
The Response time setting is used to limit current
draw when input voltage less than 1V. The
Response setting sets the bandwidth of the load to
Fast (20kHz) or as Normal (200Hz).
Parameters
Response
Normal, Fast
Panel operation
Ensure the menu is the configuration menu. See
page 173.
Configure
Previous
Menu
Protection
Go-NoGo
Other
Group
1.
Press the F2 (Other) key to enter the
other menu.
F1
F5
~
F2
Summary of Contents for PEL-2000B Series
Page 1: ...Programmable DC Electronic Load PEL 2000B Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Page 5: ...Table of Contents 5 INDEX 312...
Page 113: ...TUTORIALS 113 14 Sound Turn the sound on off for the mainframe IU Page 202...
Page 309: ...APPENDIX 309 PEL 2004B 435 4 177 532 3 195 3 481 8 554 1 533 3 583 3...
Page 310: ...PEL 2000B Series User Manual 310 PEL 2020B PEL 2030B PEL 2040B PEL 2041B 81 6 171 2 475 446 8...
Page 317: ...INDEX 317 Wiring procedure 46...