Slew rate
The slew rate is the rate at which the current will
increase to a set level. There are two slew rates:
rising slew rate & falling slew rate. In CC mode
the slew rate is defined as A/uS.
Current
Time
Level 1
Level 2
Slew Rate (rise)
A/uS
Slew Rate (fall)
A/uS
Time 1
Time 2
Slew Rate
As can be seen above, the rising and falling slew
rate need not be the same.
Go/NoGo
Using Go/NoGo, the Center, High and Low
voltage limits can be set for both Static and
Dynamic modes. A delay time of up to 1 second
can also be set.
Load
Current
Voltage
H
ig
h
Go
L
o
w
CC Mode: Static:Go/NoGo
Current
NoGo
NoGo