OPERATING DESCRIPTION
89
External Voltage Control
Voltage/
Current
Setting
(Percent
Rating)
External Voltage
100%
0%
10V
0V
To determine the Percent Rating (voltage or
current load input), use the following formula;
VorA
Rating
V
Voltage
External
Input
Load
10
Where “Rating V or A” is the rating
voltage/current of the load module.
Current Monitor
The load current input can be externally monitored
using the IMON pin of a channel control
connector. The IMON pin outputs a voltage of
0~10V to represent the input current as a
percentage (0~100%) of rating current.
Current Monitor
Current
Monitor
Output
Load Input Current
10V
0V
Rating current
0A
To determine the Current Monitor Output
(IMON) , use the following formula;
V
A
Rating
current
input
Load
IMON
10
Where “Rating A” is the rating current of the load
module.