OPERATION
193
Go/NoGo can be used in both high and low
ranges, as well as CC, CV, CR and CP Modes. The
Go/NoGo status can be read using the rear
Go/NoGo output.
A delay time can also be imposed for up to 1
second.
Note
Any Go/NoGo configuration only applies to the
current channel with the same mode and range.
Parameters
Entry Mode Value
Percent
CC Mode
CR Mode
High: V
Low: V
High: %
Low: %
Center: V
CV Mode
CP Mode
High: A
Low: A
High: %
Low: %
Center: A
Delay Time 0.0~1.0 seconds
SPEC Test ON/OFF
Panel Operation
Ensure the menu is in the Configuration menu.
See page173.
Configure
Previous
Menu
Protection
Go-NoGo
Other
Group
Choose Absolute/
Percentage limits
1.
Press (F3) Go-NoGo to access the
Go/NoGo menu.
F1
F5
~
F3
2.
Use the selector knob to edit
Entry Mode.
3.
Choose Value for absolute limits or Percent for
percentage offset limits.