Voltage Monitor
The input voltage, like the load input current can
be externally monitored with the channel control
connectors. The VMON pin of the channel control
connector outputs a voltage of 0~10V to represent
the load input voltage as a percentage (0~100%) of
the rating voltage.
Voltage Monitor
Voltage
Monitor
Output
Load Input
Voltage
10V
0V
Rating voltage
0V
To determine the Voltage Monitor Output
(VMON) , use the following formula;
V
V
Rating
voltage
input
Load
VMON
10
Where “Rating V” is the rating voltage of the load
module.
Turning on the
Load
A load is turned on when Load On input is set to
On (active low). A load is turned off when the
Load On input is set to Off (active high).
When a load is turned on from the channel control
interface, the load can be turned off from the
mainframe, local module and via remote control.
However the opposite is not true; when a load is
turned off using the channel control interface, the
load cannot be turned on via the mainframe, local
module or via remote control.
For connections and configurations, see pages 55
and 271 respectively.
Summary of Contents for PEL-2000B Series
Page 1: ...Programmable DC Electronic Load PEL 2000B Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Page 5: ...Table of Contents 5 INDEX 312...
Page 113: ...TUTORIALS 113 14 Sound Turn the sound on off for the mainframe IU Page 202...
Page 309: ...APPENDIX 309 PEL 2004B 435 4 177 532 3 195 3 481 8 554 1 533 3 583 3...
Page 310: ...PEL 2000B Series User Manual 310 PEL 2020B PEL 2030B PEL 2040B PEL 2041B 81 6 171 2 475 446 8...
Page 317: ...INDEX 317 Wiring procedure 46...