OPERATION
135
Level1
Ω
100.000
3.
Press the Selector knob or Enter
to confirm selection.
ENTER
or
4.
Repeat steps 1-3 for the remaining parameters.
Note
Level1 & Level2 can be set for both High and Low
Range.
*Press Shift to toggle between coarse and fine
adjustment when editing the Level1 and Level2
parameters. See page 188 for details.
Select CR Static Mode
Background
Constant Resistance mode can be set to dynamic
or static mode. Static mode is for manually
varying the load for single channel load modules
or to set a static load on dual channel modules.
Ensure the menu is in CR Mode. See page 131.
Configure
Configure
Mode
CC
Dynamic
Range
High
Panel Operation
1.
Press the F3 key until Static mode is
selected.
Dynamic
Static
F1
F5
~
F3
Note
Changing from static to dynamic mode will only affect
the current (active) channel.