PEL-2000B Series User Manual
214
Configure
Other
Other Setting
High Resolution
System Mode
Von Latch Clear
Measure Period
Jog Shuttle Control
ON
0
Auto
200ms
OFF
Load
Interface
LOAD
USB
Time Set
System
Info
05/04/22
16 : 50
3.
Use the Selector knob to highlight
System Mode.
4.
Press the Selector knob to edit
System Mode, then turn to
change from
0
to
1
and vice versa.
System Mode
0
5.
Press the Selector knob or Enter
to confirm selection.
ENTER
or
Adjusting the Von Latch Clear
Background
Auto: Load starts when the terminal voltage of
module is higher than Von value. The system
stops loading when the terminal voltage of
module is close to 0V for more than 25ms and
system is under the state of detecting Von again.
Manual: The load starts when the terminal voltage
of module exceeds the Von setting value. Loading
keep going even if the terminal voltage of module
close to 0V.