GETTING STARTED
45
Current
Voltage
time
time
∆I
∆T
E
Min. V
The diagram above shows how changes in current
can affect voltage.
Limiting Load line
inductance
Load line inductance can be reduced by ensuring
load wires are as short as possible and by twisting
positive and negative load wires together. Current
change can be limited by limiting the slew rate
when switching.
“Twisted pair” will be shown on any connection
diagram where the load wires should be twisted
together.
DUT
-
+
Programable
Electronic
Load
-
+
Twisted
Pair
DUT
-
+
Programable
Electronic
Load
-
+
DUT
-
+
Programable
Electronic
Load
-
+