OPERATING DESCRIPTION
67
CV + CC Mode
Load Input
Voltage
Input Current
Current Limit
Response Speed
Response speed can be set to fast, slow1, slow2 or
slow3. Fast response and slow1 (slow2/slow3)
response are determined by the load module
specifications. Slow response speeds are suitable
for large loads as quick current changes will
induce induction which can cause large voltage
drops. The PEL-2000B series will try to rectify any
voltage drops. However if voltage drops are too
large, they may cause the load to go into
oscillation. Large voltage drops caused by line
voltage induction may damage the machine.
Go/NoGo
Go/NoGo testing can be with either current
(Ampere) values (High, Low) or percentage values
(Center, High %, Low %). A delay time of up to 1
second can also be set.
CV Mode: GO/NOGO
Voltage
Load
Current
High
Low
Load
Voltage
Go
NoGo
NoGo