PEL-2000B Series User Manual
160
Note
Add Point will insert a new point directly after the
current point. The value of the current/resistance of
the inserted point will be the average of the point
before and after. All other settings will remain
unchanged.
Delete Point
9.
To delete the current point, press
Delete Point (F2).
F1
F5
~
F2
Edit previous
point
10.
Use the selector knob to change
the current point number.
Point
001
Note
The Point number can only be changed if more points
have already been added.
Save Sequence
11.
Press Save (F3) to save the sequence.
F1
F5
~
F3
Note
The save icon will only appear after a change has
been made in the menu.
Create Sequence Loop
Background
Sequences can be looped a number of times. The
loops can be started at any point in the sequence.
The Start of Loop function determines which
point will start each repeating loop.
The On End of Seq. function will hold the load
current (of the selected sequence) to a designated
value until all the other sequences have finished
running.
CC Vrange sets the range in CC mode for
Sequences. See page 75 for more details.
Summary of Contents for PEL-2000B Series
Page 1: ...Programmable DC Electronic Load PEL 2000B Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Page 5: ...Table of Contents 5 INDEX 312...
Page 113: ...TUTORIALS 113 14 Sound Turn the sound on off for the mainframe IU Page 202...
Page 309: ...APPENDIX 309 PEL 2004B 435 4 177 532 3 195 3 481 8 554 1 533 3 583 3...
Page 310: ...PEL 2000B Series User Manual 310 PEL 2020B PEL 2030B PEL 2040B PEL 2041B 81 6 171 2 475 446 8...
Page 317: ...INDEX 317 Wiring procedure 46...