INTERFACE
279
Connect the RS-232C to RJ45 cable to
the rear panel port: RJ45 female
connector.
RS232/
RS485
IN
OUT
Terminal
application
Invoke a terminal application such as MTTTY
(Multi-Threaded TTY).
For RS-232C, set the COM port, baud rate, stop
bit, data bit, and parity accordingly.
To check the COM port No. for RS-232C, see the
Device Manager in the PC. For Win XP, Control
panel → System → Hardware tab.
5.
Ensure the terminal application has the
following settings;
6.
Baud rate – as per PEL-2000B settings
7.
Com Port – as per PC settings (Device
Manager)
8.
Parity – None
9.
Data bits – 8
10.
Stop bits – None
Functionality
check
Run this query command via the terminal.
*idn?
This should return the Manufacturer, Model
number, Serial number, and Firmware version in
the following format.
GW, PEL-2002B/2004B, 00000001, V3.01