Module Panel
Description
All positions in ÄKTA pure must be occupied. Positions not used for core or optional modules must be
fitted with a Module Panel.
Module Panels are installed in the same way as the other optional modules and the cable inside must
be connected to the Module Panel, see Hardware installation of a module, on page 77.
When an optional module is replaced by a Module panel, the removed module has to be deselected
in the software configuration, see Edit system properties, on page 128.
Illustration
The illustration shows the Module Panels installed in the example configuration used in this manual.
ÄKTA pure User Manual 29119969 AB
85
2 The ÄKTA pure instrument
2.6 Accessories