Action
Step
In the Manual instructions dialog:
1
• Select Flow path:Injection valve.
• Select Sample pump load from the Position drop-down list.
• Click
.
Result: The Injection valve switches to Sample pump load position.
Make sure that the sample inlet tubing from the sample vessel is connected to the Sample
inlet valve.
2
In the Manual instructions dialog:
3
• Select Alarms:Alarm sample pressure.
• Set Mode as Enabled.
• Set a High alarm level that is below the maximum pressure of the Superloop.
• Click
.
In the Manual instructions dialog:
4
• Select Pumps and pressures:Sample flow.
• Set Flow rate to an appropriate value for the Superloop size, in this example 10 ml/min.
• Click
.
Result: A sample flow starts, in this example of 10 ml/min.
When the Superloop is filled with as much volume as is needed, click the End icon in the
System Control toolbar to end the run.
5
Plug the Syr port on the Injection valve with a Stop plug.
6
186
ÄKTA pure User Manual 29119969 AB
5 Operation
5.7 Sample application
5.7.2 Sample application using a Superloop
™