Action
Step
Pull off the detector and the flow cell from the monochromator. Be careful not to damage
the UV flow cell.
4
Note:
Make sure that the flow cell does not come into contact with any liquid, and that no liquid
enters the UV detector or monochromator.
Note:
While the UV detector is disconnected, protect the fiber connectors from dust or other im-
purities by mounting the rubber protective caps onto them.
Note:
Do not touch the optical fiber connectors as this will result in poor monitor performance.
If you accidentally touch the optical fiber connectors, clean them according to Clean the
optical fiber connectors, on page 327.
Pull off the UV flow cell from the detector.
5
Pull off the black protective caps from the new UV flow cell, and connect the new UV flow
cell to the detector.
6
Connect the detector, with the new flow cell connected, to the monochromator. Pull the
latch upwards to fasten the detector.
7
Connect the tubing to the new flow cell.
8
Switch on the instrument.
9
Result: The flow cell path length is automatically recognized by the monitor when a new
flow cell is connected.
326
ÄKTA pure User Manual 29119969 AB
7 Maintenance
7.8 Replacement procedures
7.8.5 Replace the UV monitor U9-M flow cell