5.8.1
Prepare Fraction collector F9-C
About this section
This section describes how to prepare and assemble Fraction collector F9-C before a run.
The fraction collector is connected to ÄKTA pure and controlled by UNICORN. Control of the fraction
collector can be achieved automatically in a method run, or manually.
CAUTION
Fire Hazard. Do not fractionate flammable liquids using Fraction collector F9-C.
When running RPC methods, or other procedures using organic solvents, collect
fractions through the outlet valve or Fraction collector F9-R.
The tray and racks can tilt slightly when not fully inserted into the fraction collector and may
harm the fractionation arm. The tilt is due to the height difference between the door and the
floor of the fraction collector. The tendency to tilt is affected by the placement and weight of
the cassettes, tubes or bottles
Note:
Prepare the fraction collector
Before starting to prepare Fraction collector F9-C, check the fractionation settings in the method to
be run. Perform the steps described below according to the settings in the method.
• Insert the Cassette tray or a rack for tubes or bottles.
• Change the System Settings in UNICORN to set the fractionation mode and other settings for
fraction collection.
How to insert a tray or a rack is shown below.
For information on how to change the System Settings before a run, see UNICORN System Control
Manual. The available System Settings are described in the ÄKTA pure User manual.
194
ÄKTA pure User Manual 29119969 AB
5 Operation
5.8 Fractionation
5.8.1 Prepare Fraction collector F9-C