9.8.6
System settings - I/O-box
The table below describes the I/O-box related system settings available for ÄKTA pure.
Description
Instruction name
Digital out X sets the value of the signal sent out by digital port
number X to either 0 or 1. The default value is 1.
Digital out X
Noise reduction analog in X filters the noise in the analog signal
in port number X.
Noise reduction analog in X
Alarm analog in X enables or disables the alarm for the analog
signal in port number X. When enabled, it sets the alarm limits for
the analog signal. If the alarm is enabled and the analog signal
falls outside the set limits, an alarm will be triggered and the
method will be paused.
Alarm analog in X
Alarm digital in X enables or disables the alarm for the signal in
digital port number X. The alarm can be triggered by either of the
signal values, 0 or 1. If the alarm is enabled and the condition set
in ‘Value’ occurs, an alarm will be triggered and the method will
be paused.
Alarm digital in X
Configure analog out X enables the user to send one of the pre-
defined signals (UV signal, conductivity, temperature, pH or con-
centration of eluent B) to the analog out port number X, and also
to set the range of that signal.
Configure analog out X
ÄKTA pure User Manual 29119969 AB
457
9 Reference information
9.8 System settings
9.8.6 System settings - I/O-box