Action
Description
Error code
The instrument restarts. If the error is recurrent,
generate a system error report and contact ser-
vice.
Fraction collector configuration
error.
33 - 35, 37 -
40, 55
Clean the cassette code reader, the cassette type
code on the cassette and the plate.
The fraction collector failed to
detect the code on the cassette
or the type of plate.
47 - 48
Clean the Drop sync sensor, see user documen-
tation for the location of the sensor. If this error
is recurrent, please contact service.
Calibration of the dropsync sen-
sor failed.
53
Close the door of the fraction collector.
Fraction collector movement is
blocked. Either the door is open
or there are communication
problems.
54
The instrument restarts. If the error is recurrent,
generate a system error report and contact ser-
vice.
Internal fraction collector error.
57 - 59
Lower the flow rate or increase the fractionation
volume.
Tube change is too fast.
63
The instrument restarts. If the error is recurrent,
generate a system error report and contact ser-
vice.
Fraction collector arm movement
is blocked.
68, 72
Fraction collector F9-R
Action
Description
Error code
See Section 8.5 Troubleshooting: Fraction collec-
tor, on page 377
High temperature
20
Clean the sensor and remove air bubbles in the
flow path.
Drop sync warning
21
Check that the tube sensor is adjusted properly.
Tube sensor error
22
Increase the fraction size. Fraction size must be
greater than 1/10 of the delay volume.
Delay queue full
23
Restart the instrument. If recurrent please con-
tact Service.
Internal instrument error
24, 26
Increase the fraction size or lower the flow rate.
Too fast tube change
25, 28
408
ÄKTA pure User Manual 29119969 AB
8 Troubleshooting
8.10 Error codes