Possible cause and action
Problem
• The air intake on the rear or on the left side of the chromatog-
raphy instrument or the sample pump is covered
Make sure that none of the air intakes on the instrument are
covered.
• Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35°C.
• Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
The internal temperature of
the pressure monitor is too
high
ÄKTA pure User Manual 29119969 AB
373
8 Troubleshooting
8.3 Troubleshooting: Monitors