9.9.7
Manual instructions - Wash settings
The table below describes the wash related manual instructions available for ÄKTA pure.
Description
Instruction name
System wash settings sets the flow rate used for System wash.
Note:
• The volume for system wash is set in the System wash instruc-
tion.
• The flow rate should not exceed 10 ml/min if narrow inlet tubing
(I.D. 0.75 mm) is used.
• Adjust the flow rate during the system wash so that the system
pressure does not exceed 2 MPa.
System wash settings
System pump wash settings sets the flow rate and wash volume
used during system pump washes.
Note:
The flow rate should not exceed 10 ml/min if narrow inlet tubing
(I.D. 0.75 mm) is used.
System pump wash settings
Sample pump wash settings sets the flow rate and wash volume
used during sample pump washes.
Sample pump wash settings
Fraction collector wash settings sets the flow rate during Frac-
tion collector wash.
Fraction collector wash settings
Mixer by-pass wash settings sets the flow rate used during Mixer
by-pass wash and defines wash volumes for mixer by-pass wash
options.
Note:
The flow rate should not exceed 10 ml/min if narrow inlet tubing
(I.D. 0.75 mm) is used.
Mixer by-pass wash settings
Loop wash settings sets the flow rate used during Loop wash.
Note:
• The volume for system wash is set in the Loop wash instruction.
• The flow rate should not exceed 10 ml/min if narrow inlet tubing
(I.D. 0.75 mm) is used.
Loop wash settings
482
ÄKTA pure User Manual 29119969 AB
9 Reference information
9.9 Manual instructions
9.9.7 Manual instructions - Wash settings
Summary of Contents for AKTApure
Page 1: ...ÄKTA pure User Manual ...
Page 2: ...Page intentionally left blank ...
Page 520: ...Page intentionally left blank ...