3.6 Connection of external modules
Introduction
The external modules are not installed in the instrument cabinet, but are connected via a UniNet-9
cable at the back of the system. it is possible to install up to six external modules with F-type connectors
and up to two external modules with D-type connectors at the same time.
Some of the external modules are also connected to the ÄKTA pure instrument by tubing. See the ap-
propriate section in Chapter 3 ÄKTA pure external modules, on page 87 for detailed instructions on
how to install this tubing, if applicable.
F-type and D-type connectors
The illustrations below show the F-type and D-type connectors.
D-type
F-type
Constraints
The table below indicates usage constraints for the different external modules.
Constraints
Connector
External module
I/O-box E9 has no constraints.
F-type
I/O-box E9
I/O-box E9, 2nd requires I/O-box E9.
F-type
I/O-box E9, 2nd
External air sensor L9 has no constraints.
F-type
External air sensor L9
External air sensor L9, 2 requires External air sensor L9.
F-type
External air sensor L9, 2
External air sensor L9, 3 requires External air sensor L9,
2.
F-type
External air sensor L9, 3
External air sensor L9, 4 requires External air sensor L9,
3.
F-type
External air sensor L9, 4
Fraction collector F9-C requires an Outlet valve.
D-type
Fraction collector F9-C
Fraction collector F9-R requires an Outlet valve.
F-type
Fraction collector F9-R
122
ÄKTA pure User Manual 29119969 AB
3 ÄKTA pure external modules
3.6 Connection of external modules