Possible cause and action
Problem
• The door of the fraction collector has been opened between
two runs. When the instrument is in state Ready and the door
is opened and closed, the Fraction collector content is reset.
If fractionation will continue after the last fraction of the first run,
do not open the door of the fraction collector between two runs.
The fractionation starts in
the first row again
• The combination of high flow rates and liquids with low surface
tension might lead to spillage in the Fraction collector.
If possible use Cassette positions 5 and 6. Lower the flow rate
when using liquids with low surface tension.
Spillage between fractions
Fraction collector F9-R
Possible cause and action
Problem
• The cable between the fraction collector and the ÄKTA pure
instrument is not connected
Make sure that the cable is connected.
• Wrong node ID
Check the Node ID of the fraction collector. If necessary, change
the Node ID. See Section 9.16 Node IDs, on page 503.
The fraction collector can-
not be found by the instru-
ment
• Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35°C.
• Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact service.
The internal temperature of
the fraction collector is too
high
ÄKTA pure User Manual 29119969 AB
385
8 Troubleshooting
8.5 Troubleshooting: Fraction collector