Description
Sampling
frequency
Range
Curve
The pH signal.
1 Hz
0.00 – 14.00
pH
The UV absorbance signal of the
U9-L monitor.
10 Hz
-6000.000 – 6000.000 mAU
UV
The UV absorbance signal of the
second U9-L monitor.
10 Hz
-6000.000 – 6000.000 mAU
UV 2nd
The system pressure signal (at the
system pumps).
10 Hz
-1.00 – 20.00 MPa (ÄKTA
pure 25)
System pressure
-1.00 – 5.00 MPa (ÄKTA pure
150)
The sample pressure signal (at the
sample pump).
10 Hz
-1.00 – 10.00 MPa (ÄKTA
pure 25)
Sample pressure
-1.00 – 5.00 MPa (ÄKTA pure
150)
The post-column pressure signal.
1 Hz
-1.00 – 20.00 MPa (ÄKTA
pure 25)
PostC pressure
-1.00 – 5.00 MPa (ÄKTA pure
150)
The temperature signal (in the con-
ductivity flow cell).
0.5 Hz
0.0 – 99.0 °C
Cond temp
The I/O-box analog input signals.
10 Hz
-2000.0 – 2000.0 mV
Analog in 1, to
Analog in 4
The nominal cell path length of the
U9-L monitor.
1 Hz
0.2, 0.5 cm
UV cell path
length
The nominal cell path length of the
U9-M monitor.
1 Hz
0.05, 0.2, 1.0 cm
UV cell path
length
The nominal cell path length of the
2nd U9-L monitor.
1 Hz
0.2, 0.5 cm
UV cell path
length 2nd
The I/O-box digital input signals.
10 Hz
0, 1
Digital in 1, to
Digital in 8
The I/O-box digital output signals.
10 Hz
0, 1
Digital out 1, to
Digital out 8
492
ÄKTA pure User Manual 29119969 AB
9 Reference information
9.11 Available Curves