Waste tubing overview
The table below lists the waste tubing of the instrument and where it is located. Make sure that the
waste tubing is connected to the correct positions on the modules.
Location of tubing
Tubing connections
Module
Front of the ÄKTA pure instrument.
Waste ports W1 and W2
Injection valve
Front of the ÄKTA pure instrument.
Waste port W3
pH valve (optional)
Front of the ÄKTA pure instrument.
Waste port W
Outlet valve (optional)
Rear of the ÄKTA pure instrument.
Drainage hole of the Buffer
tray
Buffer tray (Rescue drainage)
Rear of the fraction collector.
Fraction collector waste outlet
Fraction collector F9-C
(optional)
Prepare waste tubing
Follow the instructions below to prepare the waste tubing.
CAUTION
Fasten the waste tubing. During operation at high pressure the ÄKTA pure instrument
may release bursts of liquid in the waste tubing. Securely fasten all waste tubing to
the ÄKTA pure instrument and to the waste vessel.
CAUTION
Make sure that the waste vessel will hold all the produced volume of the run. For
ÄKTA pure, a suitable waste vessel should typically have a volume of 2 to 10 liters.
NOTICE
The maximum level of the waste vessel must be lower than the bottom of the ÄKTA
pure instrument.
ÄKTA pure User Manual 29119969 AB
155
5 Operation
5.2 Prepare the flow path