6.7 Fraction Collector F9-R Test
Method description
The Fraction collector F9-R test checks the functionality of Fraction collector F9-R.
The method run takes approximately 3 minutes.
Required configuration
A correctly installed Outlet valve, and a correctly installed Fraction collector F9-R are required to run
the test.
Required material
The following materials are required:
• Distilled water
• Syringe, 25-30 ml
• 7 tubes for collecting the fractions.
Prepare the test
Follow the instructions below to prepare the system before method start.
Action
Step
Direct outlet tubing W to a waste container.
1
Place inlet tubing A1 into distilled water.
2
Prime inlet A1 and purge System pump A. See Section 5.4.1 System pumps, on page 162.
3
Place 7 tubes in the Fraction collector, in positions 1 to 7.
4
In the System Control module, select System:Settings.
5
Result: The System Settings dialog opens.
In the System Settings dialog:
6
• Select Fraction collector:Fractionation settings.
• In the Drop sync field, select On.
• Click OK.
Run and evaluate the test
Follow the instructions described in Section 6.1 General performance test actions, on page 226 to start
and run the performance test.
240
ÄKTA pure User Manual 29119969 AB
6 Performance tests
6.7 Fraction Collector F9-R Test