9 Measurement
FLUXUS F60*
9.2 Measurement settings
99
UMFLUXUS_F60xV5-0EN, 2017-10-16
9.2.4
Activation of an analog output
• Select the channel for which an output is to be activated in the program branch
Output
Options
.
• Press ENTER.
• Press ENTER until
Current Loop
is displayed. Select
yes
to activate the output.
• Press ENTER.
Output Options
for Channel …
ENTER until
Current Loop
Meas.Values
Zero-Scale Val.
Full-Scale Val.
Test output
Error-val. delay
see annex A, p. 224
Notice!
An output can only be activated in the program branch
Output Options
if it has
previously been installed.
Output Options\for Channel A:
Output Options\...\Current Loop