18 SuperUser mode
FLUXUS F60*
18.3 Definition of the flow parameters
177
UMFLUXUS_F60xV5-0EN, 2017-10-16
18.3.2
Correction of the flow velocity
After the profile bounds have been defined, it is possible to define a correction of the flow
velocity:
v
cor
= m · v + n
where
All quantities derived from the flow velocity will be calculated with the corrected flow ve-
locity. The correction data will be transmitted to the PC or printer during the online or
offline transmission.
• Select
on
to define the correction data,
off
to work without correction of the flow veloc-
ity and return to the menu item
SYSTEM settings
.
• If
on
is selected, enter the slope. The input
0.0
deactivates the correction.
• Press ENTER.
• Enter the offset. Enter zero to work without offset.
• Press ENTER.
v
–
measured flow velocity
m
–
slope, range: -2.0...+2.0
n
–
offset, range: -12.7…+12.7 cm/s
v
cor
–
corrected flow velocity
Notice!
It will not be displayed that the correction of the flow velocity is active during the
measurement.
Special Funct.\...\Calibration\...\Calibration
Special Funct.\...\Calibration\...\Calibration\Slope
Special Funct.\...\Calibration\...\Calibration\Offset