13 Outputs
FLUXUS F60*
13.1 Configuration of an output using the adapter for the active current input
121
UMFLUXUS_F60xV5-0EN, 2017-10-16
When configuring an analog output, the output range is defined.
• Select
other range
to enter the output range manually.
• Press ENTER.
• Enter the value
4 mA
.
• Press ENTER.
• Enter the value
22 mA
.
• Press ENTER.
• Select the list item
Other value
for the error output.
• Press ENTER.
• Enter the error value
22 mA
.
• Press ENTER.
Activation of the output
• Select the channel for which an output is to be activated in the program branch
Output
Options
.
• Press ENTER.
• Press ENTER until
Current Loop
is displayed. Select
yes
to activate the output.
• Press ENTER.
After an analog output has been activated in the program branch
Output Options
, the
measuring range of the source item has to be entered.
• Select
absolut
.
• Press ENTER.
Special Funct.\SYSTEM settings\Proc. outputs\...\I1 Output range
Special Funct.\SYSTEM settings\Proc. outputs\...\I1 Output MIN
Special Funct.\SYSTEM settings\Proc. outputs\...\I1 Output MAX
Special Funct.\SYSTEM settings\Proc. outputs\...\Error-value
Output Options\...\Current Loop
Output Options\...\Meas.Values