13 Outputs
13.1 Configuration of an output using the adapter for the active current input
FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
120
13
Outputs
13.1
Configuration of an output using the adapter for the active
current input
Installation of the output
• Select
Special Funct.\SYSTEM settings\Proc. outputs
.
• Press ENTER.
• Select the output
Current I1
.
• Press ENTER.
A tick (
) after a list item indicates that this output has already been installed.
This display will be indicated if the output has not been installed yet.
• Select
yes
to install or reconfigure the output.
• Press ENTER.
• Select the measuring channel to be assigned to the output as the source channel in the
scroll list.
• Press ENTER.
• Select
Miscellaneous
in the scroll list.
• Press ENTER.
• Select
Signal
in the scroll list.
• Press ENTER.
Special Funct.\SYSTEM settings\Proc. outputs.
Special Funct.\SYSTEM settings\Proc. outputs\Install Output
Special Funct.\SYSTEM settings\Proc. outputs\I1 enable
Special Funct.\SYSTEM settings\Proc. outputs\...\I1 Source chan.
Special Funct.\SYSTEM settings\Proc. outputs\...\Source item
Special Funct.\SYSTEM settings\Proc. outputs\...\Miscellaneous