14 Inputs
14.4 Temperature correction
FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
134
14.4.1
Activation/deactivation of the temperature correction
The temperature correction can be activated/deactivated in the program branch
Special Funct.\SYSTEM settings\Dialogs/Menus
.
• Select
on
to activate the temperature correction,
off
to deactivate it.
• Press ENTER.
14.4.2
Input of the temperature correction
During the flow transducer positioning, the correction values will be requested for each
input which has been activated and where the temperature can be measured.
• Enter the offset for the temperature input.
• Press ENTER.
Special Funct.\SYSTEM settings\Dialogs/Menus
Special Funct.\SYSTEM settings\Dialogs/Menus\Tx Corr.Offset
Notice!
If
off
is selected, the temperature correction will be deactivated for all inputs. How-
ever, the entered correction values for each temperature input will be stored and dis-
played again when the temperature correction is activated again.
T1 Corr.Offset\0.3 C
Notice!
Only measured temperatures can be corrected.
In order to adjust the zero point, the same reference temperature is measured with
both temperature probes. The difference of both measured temperatures is entered
as the offset for one of the temperature inputs. The difference can also be distrib-
uted between the offsets of both channels.
The display of the temperature difference T1-T2 does not indicate if one or both tem-
peratures are constant or if the values have been corrected.