19 Settings
19.1 Dialogs and menus
FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
186
19.1.2
Fluid pressure
The dependence of the parameters of a fluid on the pressure can be taken into account.
• Select
on
to request the fluid pressure in the program branch
Parameter
. Select
off
if
1 bar is to be used for all calculations.
• Press ENTER.
19.1.3
Measuring point number
• Select (
1234
) if the measuring point is to be identified only by numbers, point and hy-
phen.
• Select (
↑↓←
→
) if the measuring point is to be designated with ASCII characters.
• Press ENTER.
Notice!
The edition of the pipe circumference is only temporarily. When the transmitter
switches back to the display of the pipe circumference (internal recalculation), slight
rounding errors may occur.
Example
entered pipe circumference: 100 mm
displayed outer pipe diameter: 31.8 mm
When the transmitter switches back to the display of the pipe circumference, 99.9 mm
will be displayed.
Special Funct.\...\Dialogs/Menus\Fluid pressure
Notice!
For documentation purposes, it is useful to enter the fluid pressure, even if the trans-
mitter contains no pressure-dependent characteristic curves.
Special Funct.\...\Dialogs/Menus\Meas.Point No.: