18 SuperUser mode
18.1 Activation/deactivation
FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
174
18
SuperUser mode
The SuperUser mode offers the possibility of an advanced diagnostic of signals and
measured values as well as the definition of additional parameters adapted to the mea-
suring point, in order to achieve better measured values or for experimental work. Spe-
cial features of the SuperUser mode are:
• Default settings will not be used.
• There are no plausibility tests during the parameter entry.
• It is not checked whether the entered parameters are within the limits given by physical
laws and technical data.
• The cut-off flow is not activated.
• A number of sound paths has to be entered.
Some menu items that are not visible in the normal mode are displayed.
18.1
Activation/deactivation
• Enter the HotCode
071049
immediately after the transmitter has been switched on.
It is displayed that the SuperUser mode is activated.
• Press ENTER. The main menu will be displayed.
The SuperUser mode is deactivated by switching off the transmitter.
Notice!
The SuperUser mode is intended for experienced users with advanced application
knowledge. The parameters can affect the normal measuring mode and lead to
wrong measured values or to a measurement failure when a new measuring point is
set up.
SUPERUSER MODE\IS ACTIVE NOW
Notice!
Some of the defined parameters are still active after the deactivation of the SuperUser
mode.