13 Outputs
13.2 Installation of a binary output
FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
124
Function test of the binary output
The function of the output can now be tested.
• Connect an external measuring instrument to the terminals of the output.
• Select
Reed-Relay OFF
or
Open collect OFF
in the scroll list
Output Test
to
test the de-energized state of the output.
• Press ENTER. Measure the resistance at the output. The value has to be high ohmic.
• Select
yes
to repeat the test,
no
to return to the menu item
SYSTEM settings
.
• Press ENTER.
• Select
Reed-Relay ON
or
Open collect. ON
in the scroll list
Output Test
to test
the energized state of the output.
• Press ENTER. Measure the resistance at the output. The value has to be low ohmic.
• Select
yes
to repeat the test,
no
to return to the menu item
SYSTEM settings
.
• Press ENTER.
Tab. 13.1: Configuration of the binary outputs
source item
list item
output
Limit
R1
limit message (
Alarm Output R1
)
R2
limit message (
Alarm Output R2
)
R3
limit message (
Alarm Output R3
)
Impuls
from abs(x)
pulse without sign consideration
from x > 0
pulse for positive measured values of the volumetric flow rate
from x < 0
pulse for negative measured values for the volumetric flow rate
Special Funct.\SYSTEM settings\Proc. outputs\...\B1 Output Test\
Reed-Relay OFF
Special Funct.\...\B1 Output Test\B1=ON\Again?
Special Funct.\SYSTEM settings\Proc. outputs\...\B1 Output Test\
Reed-Relay ON
Special Funct.\...\B1 Output Test\B1=ON\Again?