17 Advanced functions
FLUXUS F60*
17.3 HybridTrek mode
153
UMFLUXUS_F60xV5-0EN, 2017-10-16
• If
no
is selected, the NoiseTrek mode can only be activated and deactivated manually
during the measurement.
• Select
yes
to activate the automatic toggling between the TransitTime and the NoiseTrek
mode. If
yes
is selected, the NoiseTrek mode can also be activated and deactivated
manually during the measurement.
• Press ENTER.
This display will only be indicated if the NoiseTrek mode is enabled and the NoiseTrek
parallel beam mode has not been activated yet.
If the automatic toggling between the TransitTime and NoiseTrek mode is activated, the
toggling parameters have to be configured.
• Enter the time after which the transmitter has to toggle to the NoiseTrek mode if there
are no valid measured values in the TransitTime mode. If zero is entered, the transmit-
ter does not toggle to the NoiseTrek mode.
• Press ENTER.
• Enter the time after which the transmitter has to toggle from the NoiseTrek mode to the
TransitTime mode if there are no valid measured values. If zero is entered, the transmit-
ter does not toggle to the TransitTime mode.
If there are valid measured values in the NoiseTrek mode, the transmitter can periodically
toggle to the TransitTime mode in order to check whether a measurement in the TransitTime
mode is possible again. The time interval and the duration of the checking are set as follows:
• Enter the time after which the transmitter has to toggle to the TransitTime mode. If zero
is entered, the transmitter does not toggle to the TransitTime mode.
• Press ENTER.
Special Funct.\SYSTEM settings\Measuring\Enable NoiseTrek\
Auto NoiseTrek
Special Funct.\...\Auto NoiseTrek\TT-Failed |After
NoiseTrek
Special Funct.\...\Auto NoiseTrek\NT-Failed |After
TransTime
Special Funct.\...\Auto NoiseTrek\NT-Ok,but |Each check TT