19 Settings
19.3 Working with parameter records
FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
192
19.3
Working with parameter records
19.3.1
Introduction
Parameter records are data sets that contain all information necessary to perform a cer-
tain measurement task:
• pipe parameters
• transducer parameters
• fluid parameters
• output options
Working with parameter records will make repeated measurement tasks easier and fast-
er. The transmitter can store max. 14 parameter records.
The parameters have first to be entered in the program branch
Parameter
. Afterwards,
they can be stored as parameter record.
• Select
Special Funct.\Store Curr.Rec.
• Press ENTER.
The error message
NO DATA!
will be displayed if no complete parameter record is avail-
able. Storing is impossible.
• Enter the missing parameters in the program branch
Parameter
.
14 parameter records (
Par.Record
01
...
Par.Record
14
) can be stored.
• Select a parameter record.
• Press ENTER.
If parameters are already stored in the selected parameter record, they can be overwritten.
• Select
yes
to overwrite the parameters, or
no
to select another parameter record.
• Press ENTER.
Notice!
No parameter records are stored in the delivery state. Parameter records are
entered manually.
Special Funct.\Store Curr.Rec.
Special Funct.\Store Curr.Rec.\Store Par. To:
Special Funct.\Store Curr.Rec.\Store Par. To:\Overwrite