16 Data transmission
FLUXUS F60*
16.3 Data format
143
UMFLUXUS_F60xV5-0EN, 2017-10-16
• Select the serial interface for the offline transmission of data.
This display will only be indicated if the transmitter has an RS485 interface.
16.3
Data format
• Select
Special Funct.\SYSTEM settings\serial transmis.
.
SER:kill
spaces
.
• Select
on
if the space characters are not to be transmitted.
• Press ENTER.
The file size will be considerably smaller (shorter transmission time).
• Select the decimal marker to be used for floating-point numbers (point or comma).
• Press ENTER.
This setting depends on the setting of the operating system of the PC.
• Select the character to be used to separate columns (semicolon or tabulator).
• Press ENTER.
Special Funct.\SYSTEM settings\serial transmis.\Send Offline via
Special Funct.\SYSTEM settings\serial transmis.\SER:kill spaces
Special Funct.\SYSTEM settings\serial transmis.\SER:decimalpoint
Special Funct.\SYSTEM settings\serial transmis.\SER:col-separat.