17 Advanced functions
FLUXUS F60*
17.10 Modification of the limit for the inner pipe diameter
165
UMFLUXUS_F60xV5-0EN, 2017-10-16
17.10
Modification of the limit for the inner pipe diameter
It is possible to modify the lower limit of the inner pipe diameter for a given transducer
type.
• Enter the HotCode
071001
immediately after the transmitter has been switched on.
• Enter the lower limit of the inner pipe diameter of the displayed transducer type.
• Press ENTER to select the next transducer type.
17.11
Activation of a binary output as alarm output
• Select the channel for which the output is to be activated in the program branch
Output Options
.
• Press ENTER.
This display will not be indicated if the transmitter has only one measuring channel.
• Press ENTER until
Alarm Output
is displayed. Select
yes
to activate the alarm output.
• Press ENTER.
A max. of 3 independently operating alarm outputs R1, R2, R3 can be configured per
channel. The alarm outputs can be used to output information on the current measure-
ment or to start and stop pumps, motors, etc.
Notice!
If a transducer is used below its recommended inner pipe diameter, a measurement
might be impossible.
Notice!
The menu item
Alarm Output
will only be displayed in the program branch
Output
Options
if a binary output has been installed, see section 13.2.
Output Options\for Channel A:
Output Options\...\Alarm Output
DNmin Q-Sensor
15
mm