20 Wall thickness measurement (optional)
FLUXUS F60*
20.4 Measurement
203
UMFLUXUS_F60xV5-0EN, 2017-10-16
20.4.3
Further information concerning the measurement
• Press key
to obtain information on the measuring signal.
This message will be displayed if the measuring signal is sufficient. The LED of the chan-
nel lights green.
This message will be displayed if the measuring signal is not sufficient (# = number). The
LED of the measuring channel lights red.
• Press key
again. The bar graph of the signal quality (
Q=
) will be displayed.
If the signal is not sufficient for measurement,
UNDEF
will be displayed. The LED of the
measuring channel lights red.
• Shift the wall thickness probe slightly on the pipe until the LED of the measuring chan-
nel lights green.
• Press key
to display the transit time of the signal.
SIGNAL IS GOOD
3.51
mm
ERRORSIGNAL
#
mm
Q=
■■■■■■
#
3.51
mm
Wall Thickness
LZ=
186
ns