14 Inputs
14.3 Activation of the inputs
FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
132
14.3
Activation of the inputs
The activation of the inputs in the program branch
Output Options
will only be dis-
played if the transmitter has inputs of the corresponding type and they have been as-
signed to a measuring channel.
14.3.1
Activation of the temperature inputs
The temperature inputs have to be activated in case the measured temperatures are to
be displayed, stored and/or transmitted or if the measured temperature is to be used for
the interpolation of the viscosity and the density of the fluid.
• Select the channel for which the output is to be activated in the program branch
Output
Options
.
• Press ENTER.
This display will not be indicated if the transmitter has only one measuring channel.
• Select in the program branch
Output Options
the channel for which a temperature
input has to be activated. The temperature inputs assigned to the channel will be dis-
played one after another.
• Select
yes
for the temperature inputs that are to be activated.
• Press ENTER.
Notice!
If
Heatflow
is selected as the physical quantity, the corresponding temperature
inputs will be activated automatically. The steps described below are only necessary
if the measured temperatures are to be displayed or transmitted.
Output Options\for Channel A:
Output Options\...\T1:Temperature
Notice!
The total number of measured values that can be stored will be reduced if a tem-
perature input is activated.