17 Advanced functions
FLUXUS F60*
17.11 Activation of a binary output as alarm output
167
UMFLUXUS_F60xV5-0EN, 2017-10-16
17.11.2 Setting the limits
If the switching condition
MIN
or
MAX
is selected in the scroll list
func
, the limit of the out-
put has to be defined:
• Select the physical quantity to be used for the comparison in the scroll list
Input
. The
following list items are available for the alarm output R1:
– selected physical quantity
– signal amplitude
– sound speed of the fluid
• Press ENTER.
The current physical quantity is automatically set for the alarm outputs R2 and R3.
If the switching condition
MAX
is selected in the scroll list
func
, the limit of the output has
to be defined:
• Enter the upper limit.
• Press ENTER.
The alarm will switch if the measured value exceeds the limit.
If the switching condition
MIN
is selected in the scroll list
func
:
• Enter the lower limit.
• Press ENTER.
The alarm will switch if the measured value falls below the limit.
Notice!
If no measurement is made, all alarms will be de-energized, independently of the
programmed switching function.
R1 Input:\Mass Flow
R1 Input:\Function: MAX\High Limit:
R1 Input:\Function: MIN\Low Limit: