
19 Settings
19.1 Dialogs and menus
FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
188
19.1.7
Error delay
The error delay is the time after which an error value will be sent to an output if no valid
measured values are available.
• Select
edit
to enter an error delay. Select
damping
if the damping factor is to be used
as the error delay.
• Press ENTER.
19.1.8
Alarm state indication
• Select
on
to display the alarm state during the measurement.
• Press ENTER.
19.1.9
Units of measurement
It is possible to set the units of measurement for the length, temperature, pressure, den-
sity and kinematic viscosity and sound speed.
• Select a unit of measurement for all quantities.
• Press ENTER after each selection.
19.1.10 Setting relating the fluid pressure
It is possible to set whether the absolute or the relative pressure is to be used:
• Select
on
or
off
.
• Press ENTER.
If
on
is selected, the absolute pressure p
a
will be displayed/input/transmitted.
If
off
is selected, the relative pressure p
g
will be displayed/input/transmitted.
p
g
= p
a
- 1.01 bar
The pressure and its unit of measurement will, e.g., be displayed in the program branch
Parameter
. It is followed by the selected pressure, indicated in parentheses:
a – absolute pressure
g – relative pressure
Special Funct.\...\Dialogs/Menus\Error-val. delay
Special Funct.\...\Dialogs/Menus\SHOW RELAIS STAT
Special Funct.\...\Dialogs/Menus\Pressure absolut
Fluid pressure
1.00 bar(a)