9 Measurement
9.2 Measurement settings
FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
100
Measuring range
After an analog output has been activated in the program branch
Output Options
, the
measuring range of the source item has to be entered.
• Select
sign
if the sign of the measured values is to be considered for the output.
• Select
absolut
if the sign of the measured values is not to be considered for the out-
put.
• Press ENTER.
• Enter the lowest expected measured value. The unit of measurement of the source item
will be displayed.
Zero-Scale Val.
is the value assigned to the value
Output MIN
of the output range.
• Press ENTER.
• Enter the highest expected measured value. The unit of measurement of the source
item will be displayed.
Full-Scale Val.
is the value assigned to the value
Output MAX
of the output range.
• Press ENTER.
Output Options\...\Meas.Values
Output Options\...\Zero-Scale Val.
Output Options\...\Full-Scale Val.
Example
output: current output
output range: 4…20 mA
Zero-Scale Val.
: 0 m³/h
Full-Scale Val.
: 300 m³/h
volumetric flow rate = 0 m³/h, corresponds to 4 mA
volumetric flow rate = 300 m³/h, corresponds to 20 mA