7 Connection
7.2 FLUXUS *608
FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
64
Charging the battery
Connect the power supply to the transmitter, see Fig. 7.17 or Fig. 7.18. Switch on the
transmitter. The charging starts automatically. The LED "BATTERY" flashes green while
charging. The max. charging time is approx. 8 h.
During the charging process, the ambient temperature should be in the range of
0...45 °C.
A measurement can be made during the charging. Charging will be stopped automatical-
ly when the battery is fully charged. The LED "BATTERY" lights green then.
Storage of the battery
The battery remains in the transmitter. After storage, the transmitter can immediately be
operated with the battery.
• battery charge: > 30 %
• storing temperature: 12...25 °C
Notice!
The battery will only be charged if the transmitter is switched on.
Fig. 7.17:
Connection of the power supply unit to the transmitter FLUXUS *608**-A2
1 – power supply unit/battery charging unit
Fig. 7.18:
Connection of the power supply unit to the transmitter FLUXUS *608**-F2
1 – power supply unit/battery charging unit
CH A
COMM
Input
Output
CH B
DC-IN
1
1
CH A
COMM
CH B
DC-IN