
18 SuperUser mode
18.11 Display of the last valid measured value
FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
184
18.11
Display of the last valid measured value
If the signal is not sufficient for a measurement,
UNDEF
is normally displayed. Instead of
UNDEF
, it is also possible to display the last valid measured value.
• Select
Special Funct.\SYSTEM settings\Measuring\Miscellaneous
.
• Press ENTER until the menu item
Keep display val
is displayed.
• Select
on
to activate the display of the last valid measured value,
off
to deactivate it.
• Press ENTER.
18.12
Displays during the measurement
Besides the normal information, see section 9.4, the following parameters can be dis-
played during the measurement in the SuperUser mode:
Special Funct.\SYSTEM settings\Measuring\Miscellaneous\Keep display val
display
explanation
t=
transit time of the measuring signal
c=
sound speed
REYNOLD=
Reynolds number
VARI A=
standard deviation of the signal amplitude
VARI T=
standard deviation of the transit time of the measuring signal
dt-norm=
transit time difference standardized to the transducer frequency
density of the fluid