14 Inputs
FLUXUS F60*
14.2 Assignment of other inputs to the measuring channels
131
UMFLUXUS_F60xV5-0EN, 2017-10-16
• Select the temperature probe.
• Press ENTER.
• If necessary, select the temperature probe for
Input
T2
...
T4
accordingly.
14.2
Assignment of other inputs to the measuring channels
• Select
Special Funct.\SYSTEM settings\Proc. inputs
.
• Press ENTER.
• Select the list item
Link other inp.
• Press ENTER.
• Select the first input to be assigned to measuring channel A. Only the installed inputs
are displayed in the scroll list.
• Select the list item
No measuring
if no input is to be assigned to measuring channel A.
• Press ENTER.
Select the list items for
ext.Input(2)
...
(4)
of the measuring channel A and the other
activated channels accordingly.
Special Funct.\SYSTEM settings\...\Input T1
Special Funct.\SYSTEM settings\Proc. inputs
Special Funct.\SYSTEM settings\Proc. inputs\Link other inp.
Special Funct.\SYSTEM settings\...\A:ext.Input(1)\Input I1
Notice!
The configuration of a measuring channel will be stored when the next channel is
selected. The configuration dialog for a channel has to be completed in order to
store the changes.