15 Data logger
FLUXUS F60*
15.2 Setting the storage rate
137
UMFLUXUS_F60xV5-0EN, 2017-10-16
15.2
Setting the storage rate
The storage rate is the frequency to transmit or store measured values. The storage rate
is set separately for each measuring channel.
If the storage rate is not set, the storage rate previously selected will be used.
The storage interval should be at least equal to the number of activated measuring chan-
nels, e.g., the storage interval with 2 activated measuring channels: min. 2 s, 4 s are rec-
ommended.
• Select a storage rate or
EXTRA
.
• Press ENTER.
This display will only be indicated if
Store Meas.Data
and/or
Serial Output
are
activated.
• Enter the storage rate if
EXTRA
was selected.
• Press ENTER.
15.3
Configuration of the data logger
• Select
Special Funct.\SYSTEM settings\Storing
.
• Press ENTER.
It contains the following menu items:
• ringbuffer
• storage mode
• totalizer storing
• signal amplitude storing
• fluid sound speed storing
• diagnostic values storing
• acoustic signal during storing
Output Options\...\Storage Rate
Output Options\...\Storage Rate\EXTRA
Special Funct.\SYSTEM settings\Storing