Liquiline System CA80HA
Operation
Hauser
97
connect the
Measurement active
to a binary output to identify when the measurement
actually begins. If you do not require any more measurements once the first measurement is
finished, you have to already disable the
Signal for process access
during the active
measurement.
The following information is processed at the binary inputs or output at the binary outputs:
• Binary inputs:
Signal for process access
: The analyzer can only perform activities that require sample
(measurement, calibration, cleaning) when the signal is active. The time and the sequence
of activities correspond to the settings in the setup. All activities that require sample are
delayed as long as the signal is inactive.
• Binary outputs:
•
Measurement active
signal: Indicates that a measurement is currently running. The
signal is not active in the event of a calibration or cleaning.
•
Sample required
signal: The signal is active for a configurable time before every activity
that requires sample. This makes it possible to activate an external pump or a dilution
module, for example.
1. For the measurement, under
Menu
/
Setup
/
Analyzer
/
Measurement
select
Continuous
as the start condition (analyzer measures continuously, without interruption, between
the measurements).
2. Select
Menu
/
Setup
/
Inputs
/
Binary input
x:y.
3. Configure the binary inputs as follows:
Menu/Setup/Inputs/Binary input x:y
Function
Options
Info
Binary input
Selection
On
Signal type
Selection
Static signal
Signal level
Selection
• High
• Low
Specifies the active signal level:
Low
Input signals between 0 and 5 V DC
High
Input signals between 11 and 30 V DC
1)
x:y = slot no. : input number
4. Link the binary inputs with the analyzer: select
Menu
/
Setup
/
Analyzer
/
Extended
setup
/
Signal for process access
.
5. Select measuring channel
SP1
or in the case of two-channel devices select measuring
channel
SP1
or
SP2
.
6. Assign a binary input to the selected measuring channel: select
Binary input
x:y.
7. Select
Menu
/
Setup
/
Outputs
/
Binary output
x:y.
8. Configure the binary outputs as follows: