Operation
Liquiline System CA80HA
106
Hauser
10.7 Signal outputs
10.7.1 Current outputs
Liquiline System CA80 has two analog current outputs by default.
You can configure additional current outputs with extension modules.
Setting the current output range
‣
Menu
/
Setup
/
General settings
:
0..20 mA
or
4..20 mA
.
Menu/Setup/Outputs/Current output x:y
Function
Options
Info
Current output
Selection
• Off
• On
Factory setting
Off
Use this function to activate or deactivate a
variable being output at the current output
Source of data
Selection
• None
• Connected inputs
• Controller
Factory setting
None
The sources of data on offer depend on your
device version.
The main value of the analyzer and all the
sensors and controllers connected to inputs are
available for selection.
Measured value
Selection
• None
• Depends on the
Source of
data
Factory setting
None
The measured value you can select depends on
the option selected under
Source of data
.
The list of dependent measured values is provided in the
Measured value
table, subject to the
Source of
data
Apart from measured values from connected sensors, you can also select a controller as the data source. The
best way to do so is by using the
Additional functions
menu. Here, you can select and configure the current
output for outputting the controlled variable.
Range lower value
Range of adjustment and
factory settings depend on the
Measured value
You can output the entire measuring range or
just some of it at the current output. To do so,
specify the upper and lower range values in
accordance with your requirements.
Range upper value