Liquiline System CA80HA
Diagnostics and troubleshooting
Hauser
157
No.
Message
Factory settings
Tests or remedial action
S
D
F
370
Internal voltage
F
On
On
Internal voltage outside the valid range
1.
Check supply voltage.
2.
Check inputs and outputs for short-
circuiting.
373
Electronictemp. high
M
On
Off
High electronics temperature
‣
Check ambient temperature and energy
consumption.
374
Sensor check
F
On
Off
No measurement signal from sensor
1.
Check sensor connection.
2.
Check the sensor, replace it if necessary.
380
Temperature sensor
F
On
On
Temperature sensor defective in reactor, cuvette or
sample pre-heating system
‣
Contact the Service Department and quote the
number displayed.
381
Valve defective
F
On
On
Indicated valve is defective
‣
Contact the Service Department and quote the
number displayed.
385
Heating temp. low
S
On
Off
Sample pre-heating target temperature not reached
1.
Check fuse.
2.
Check cable and connector.
386
Heating temp. high
S
On
Off
Temperature of sample pre-heating too high
1.
Monitor the temperature of the sample.
2.
Disconnect the power supply if the sample
temperature does not drop.
401
Factory reset
F
On
On
Factory reset is performed
405
Service IP active
C
Off
Off
Service switch is switched on
The device can be addressed at 192.168.1.212.
‣
Switch off the service switch to change to the
saved IP settings.
412
Writing backup
F
On
Off
‣
Wait for the write process to be finished
413
Reading backup
F
On
Off
‣
Wait.